В данной версии обновлен список взаимо-загрязняющих пар а также исправлены некоторые ошибки версии v2022.C02. Полный список изменений опубликован ниже.
Скачать новую версию ПО можно в ERBA-Cloud по ссылке.
Список изменений:
Application software ver.XL1000 v2022.C05 is released to include new feature, Improvement & to address
issues reported in previous software ver. XL1000 v2022.C02.
List of Changes in ASW ver. XL1000 v2022.C05:-
- New feature: Provision to support carry over pair with any test.
- Bug fixed: Paitent entry > Worklist > sample volume, R1 position , R2 position & R3 position column is not
visible Completely - Improvement: Increase in recovery attempts (From 3 to 6) for RGT rotation error & ARM interlock with
RGT ( Error Codes=82,92,28,48) - Bug fixed: CAPA, After Viewing the archive data, system is not connecting back to live database in the
same session of Multixl - Issue fixed: CAPA, Keep Multixl in status monitor screen. After analyzer goes into sleep mode, if user
clicks start button, Analyzer wakes up from sleep mode and shows pressure error. This is fixed by increasing
wait time. - Bug fixed: CAPA, When 2 R1 bottles and 1 R2 bottle was scheduled for chemistry in the system. Then
while batch run, after 1st R1 bottle is over it use to remove R2 bottle from system. This issue is fixed and R1
bottle is switched and utilized basis on available R2 reagent.